Chapter 7 Specifications
TestEquity 115A Temperature Chamber Page 7-1
Chapter 7Specifications
Model 115A Chamber Specifications
Temperature Range 73°C to +175°C
Control Tolerance ±0.2°C (Measured at the control sensor after stabilization)
Uniformity ±1.0°C (Variations throughout the chamber after stabilization)
Live Load Capacity @ +23°C 0°C 40°C 55°C 65°C
300 W 255 W 160 W 110 W 80 W
Cool Down Transition Time*
End Temp
Start Temp to
+23°C 0°C 40°C 55°C 65°C –73°C
+23°C ---- 5 min 20 min 30 min 40 min ultimate
+85°C 16 min 25 min 45 min 57 min 69 min ultimate
Heat Up Transition Time* 5°C per minute typical
*Note: Transition times are measured after a 1 hour soak at the respective start temperature with an empty chamber,
as indicated on the temperature controller, 23°C ambient. Measured with set point beyond the start and end
temperatures. Does not include the effect of proportional band when approaching set point. Measured at nominal
line voltage of 120V, 60 Hz.
Refrigeration and Heating System
High Stage Refrigerant R-404A (Dupont HP-62)
Low Stage Refrigerant R-508B (Dupont SUVA-95)
Compressors 1/2 HP x 1/3 HP hermetic compressors in a cascade configuration.
Condenser Air Cooled
Heater Power 600 Watts
Input Power 120 V nominal (110 to 126 VAC), 60 Hz, 1 PH
Max Current Draw 15 A, Recommended Minimum Service 20 A
Workspace Dimensions 16" W x 12" H x 14" D (1.55 cubic feet)
Outside Dimensions Bench Version 24" W x 44" H x 26" D (nominal)
Floor Version 24" W x 61" H x 26" D (nominal)
Min. Installed Clearance 6" from the left and right side, 12" from the rear
Access Ports 3" Port on left and right side (two total), Supplied with foam plugs
Weight 115A-B: Bench Version, 260 pounds. 115A-F: Floor Version, 300 pounds.
Sound Level 52 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from
the floor, in a free-standing environment)
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. This
product is designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result
in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C
(85°F) or below 16°C (60°F) ambient is not recommended.