14
Accuracy (con’t)
Resistance
Measurement Range Accuracy
400.0 Ω ± (0.8% + 0.8 Ω)
4.000 ± (0.6% + 0.004 kΩ)
40.00 ± (0.6% + 0.04 kΩ)
400.0 ± (0.6% + 0.4 kΩ)
4.000 ± (1.0% + 0.004 MΩ)
40.00 ± (2.0% + 0.04 MΩ)
Diode Tester
Test Current Open Circuit Voltage
0.25 mA typical 1.6 VDC maximum
Frequency
Function Sensitivity (Sine RMS) Range
400.0 mV 350 mV 10 Hz to 2 kHz
4.000 V 1 V 10 Hz to 5 kHz
40.00 V 32 V 10 Hz to 100 kHz
400.0 V 100 V 10 Hz to 10 kHz
600 V 500 V 10 Hz to 5 kHz
400.0 A 60 A 40 Hz to 400 Hz
Accuracy of Frequency Ranges
Display Range Accuracy
5.000 Hz ± (0.5% + 0.004 Hz)
50.00 Hz ± (0.5% + 0.04 Hz)
500.0 Hz ± (0.5% + 0.4 Hz)
5.000 kHz ± (0.5% + 0.004 kHz)
50.00 kHz ± (0.5% + 0.04 kHz)
500.0 kHz ± (0.5% + 0.4 kHz)
www.calcert.com sales@calcert.com1.800.544.2843