Operation
Alarms Menu
Thermo Electron Corporation Model 42i Low Source Instruction Manual 3-79
Ozonator Flow The Ozonator Flow screen (read only) is used to display the ozonator flow
readings. If the ozonator flow reading is 0.050 LPM (50 cc) or below, an
alarm is activated, and an alarm condition screen appears as “LOW”. If the
ozonator flow is above 0.050 LPM, the no alarm condition screen is
displayed, indicating that the flow is acceptable. Inadequate ozonator flow
will cause the ozonator to overheat, resulting in permanent damage to the
ozonator.
• In the Main Menu, choose Alarms > Ozonator Flow.
Zero and Span Check The Zero Span Check screen allows the user to view the status of the most
recent zero check and set the maximum zero check offset. The zero and
span check screens are displayed if the zero/span check option is enabled
and function the same way.
• In the Main Menu, choose Alarms > Zero or Span Check.
Max Zero and Span Offset The Max Zero Check Offset screen is used to change the maximum zero
check offset. The maximum zero and span offset screens function the same
way.
• In the Main Menu, choose Alarms > Zero or Span Check > Max Offset.
OZONATOR FLOW:
ACTUAL > 0.050 LPM
RANGE AVG DIAGS ALARM
ZERO CHECK:
ALARM: OK
RESPONSE: 0.00
>MAX OFFSET: 10.00
RANGE AVG DIAGS ALARM