Operation
Service Menu
3-90 Model 42i Low Source Instruction Manual Thermo Electron Corporation
Analog Output Calibration The Analog Output Calibration menu is a selection of 6 voltage channels
and 6 current channels (if I/O expansion board option is installed) to
calibrate, and allows the user to select the calibration action zero or span.
The analog output calibration menu is visible only when the instrument is
in service mode. For more information on the service mode, see “Service
Mode” earlier in the chapter.
• In the Main Menu, choose Service > Analog Out Calibration.
CALIBRATE AMBIENT TEMP:
CURRENTLY: 28.6
o
C
SET TO: 25.0
o
C ?
§¨
MOVE CURSOR
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
WARNING This adjustment should only be performed by an instrument
service technician.
ANALOG OUTPUT CAL:
>VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
VOLTAGE CHANNEL 5
VOLTAGE CHANNEL 6
CURRENT CHANNEL 1
ª
RANGE AVG DIAGS ALARM
ANALOG OUTPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM