Operation
Service Menu
3-94 Model 42i Low Source Instruction Manual Thermo Electron Corporation
Ozonator Safety The Ozonator Safety screen is used to turn the ozonator safety feature on
and off. If the ozonator safety is turned off, the ozonator will always be on,
even if there is no flow or the converter is not up to temperature. The
ozonator safety screen is visible only when the instrument is in service mode.
For more information on the service mode, see “Service Mode” earlier in the
chapter.
In the Main Menu, choose Service > Ozonator Safety.
•Press to toggle and set the service mode on or off.
Extended Ranges The Extended Ranges screen is used to turn the extended ranges feature on
and off. The extended ranges screen is visible only when the instrument is in
service mode. For more information on the service mode, see “Service
Mode” earlier in the chapter.
In the Main Menu, choose Service > Extended Ranges.
•Press to toggle and set the extended ranges on or off.
OZONATOR SAFETY:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
EXTENDED RANGES:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM