EasyManua.ls Logo

Thermo 49i - O3 Response Coefficient; Manual O Lamp Setting

Thermo 49i
276 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Service Menu
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 3-71
O
3
Response Coefficient The O
3
Response Coefficient screen is used to change the response of the
ozonator lamp feedback circuit. The larger the number, the more rapid the
response and the higher the noise. Smaller numbers have slower response
times and less noise. Normally, this coefficient should be set to 1.0. Valid
coefficients range from 0.5 to 1.5. The O
3
response coefficient is visible
only when the calibrator is in service mode and if the ozonator option is
installed. For more information on the service mode, see “Service Mode
earlier in this chapter.
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > O
3
Response Coefficient.
Use the and to increment or decrement the numeric
value.
Press to save the new coefficient.
Manual O
3
Lamp Setting The Manual O
3
Lamp Setting screen is used to manually control the O
3
output concentration by adjusting the ozonator lamp intensity. In this
mode, the automatic feedback control of the ozonator is defeated. This is
CELL A REFERENCE GAS:
INTENSITY: 98425. Hz
NOISE: 1.4
RANGE AVG DIAGS ALARM
O3 RESPONSE COEFFICIENT:
CURRENTLY: 1.000
SET TO: 1.250 ?
©ª INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM

Table of Contents

Related product manuals