Servicing
Removing the Measurement Bench and Lowering the Partition Panel
6-8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Removing the
Measurement Bench
and Lowering the
Partition Panel
The measurement bench can be removed and the partition panel can be
lowered to improve access to connectors and components. Refer to the
following steps when a procedure requires lowering the partition panel (see
Figure 6-3).
Figure 6-3. Removing the Measurement Bench and Lowering the Partition Panel
Equipment Required:
Philips screwdriver
1. Turn instrument OFF and unplug the power cord.
2. If the instrument is mounted in a rack, remove it from the rack.
3. Remove the cover.
Pass Through
Connectors
Partition Panel
Loosen Captive Screw
Remove Screw
Remove Screws
Remove Screw
REMOVE SCREWS
SLIDE BENCH OUT
LOWER PANEL
Remove Hidden Screw
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.