Servicing
Ozonator Lamp Replacement
6-36 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
6. Reconnect the pump cable to the AC PUMP connector on the
measurement interface board.
7. Connect a calibrated flow sensor at the SAMPLE bulkhead on the rear
panel.
8. Press to return to the Flow A Sensor Cal screen.
9. At the Flow A Sensor Cal screen, press to select Span.
The Calibrate Flow A Span screen appears.
10. Wait at least 10 seconds for the reading to stabilize, use
and to enter the flow sensor reading, and press to
save the value.
11. Install the cover.
Ozonator Lamp
Replacement
Use the following procedure to replace the ozonator lamp.
Equipment Required:
Ozonator lamp
Allen wrench, 7/64-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
Equipment Damage Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.