C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual B-33
Send: set pump off
Receive: set pump off ok
temp comp
This command reports whether temperature compensation is on or off. The
example that follows shows the temperature compensation is off.
Send: temp comp
Receive: temp comp off
set temp comp onoff
These commands turn the temperature compensation on or off. The
example that follows turns temperature compensation off.
Send: set temp comp off
Receive: set temp comp off ok
l1 conc
l2 conc
l3 conc
l4 conc
These commands report the concentration setting of each of the four
custom levels. The example that follows reports that custom level 1 is 125
ppb.
Send: l1 conc
Receive: l1 conc 0125 ppb
set l1 conc value
set l2 conc value
set l3 conc value
set l4 conc value
set l5 conc value
These commands set the concentration for each of the five custom levels.
The example that follows sets custom level 1 to 325 ppb.
Send: set l4 conc 325
Receive: set l4 conc 325 ok
level d
This command sets the instrument to level d, where d represents custom
levels 1, 2, 3, 4 or 5. The example that follows sets the custom level 1.
Send: set level 1
Receive: set level 1 ok