EasyManua.ls Logo

Thermo 49i - Hardware Configuration

Thermo 49i
276 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-Link Protocol Commands
Hardware Configuration
B-34 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific
Hardware
Configuration
contrast
This command reports the screens level of contrast. The example that
follows shows the screen contrast is 50%, according to Table B-7.
Send: contrast
Receive: contrast 5: 50%
set contrast level
This command sets the screens level of contrast, according to Table B-7.
The example that follows sets the contrast level to 50%.
Send: set contrast 5
Receive: set contrast 5 ok
date
This command reports the current date. The example that follows reports
the date as December 1, 2004.
Send: date
Receive: date 12-01-04
set date mm-dd-yy
mm = month
dd = day
yy = year
This command sets the date of the analyzers internal clock. The example
that follows sets the date to March 19, 2005.
Table B-7. Contrast Levels
Level Contrast Level
00%
1 10%
2 20%
3 30%
4 40%
5 50%
6 60%
7 70%
8 80%
9 90%
10 100%

Table of Contents