Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual D-1
Appendix D Geysitech (Bayern-Hessen)
Protocol
This appendix provides a description of the Geysitech (Bayern-Hessen or
BH) Protocol Interface and is supported both over RS-232/485 as well as
TCP/IP over Ethernet.
The Geysitech Commands that are implemented are explained in detail in
this document. The Geysitech protocol support for the iSeries enables the
user to perform the functions of reading the various concentrations and to
trigger the instrument to be in sample/zero/span mode if valid for that
instrument. This is achieved by using the supported Geysitech commands
listed below.
For details of the Model 49i Primary Standard Geysitech Protocol
specification, see the following topics:
“Serial Communication Parameters” on page D-1
“TCP Communication Parameters” on page D-2
“Instrument Address” on page D-2
“Abbreviations Used” on page D-2
“Basic Command Structure” on page D-2
“Block Checksum <BCC>” on page D-3
“Geysitech Commands” on page D-3
Serial Communication
Parameters
The following are the communication parameters that are used to configure
the serial port of the iSeries to support Geysitech protocol.
Number of Data bits : 8
Number of Stop bits : 1
Parity : None