Measure UV-Vis
Thermo Scientific NanoDrop One User Guide 149
Best practices for UV-Vis measurements
Ensure the sample absorbance is within the instrument’s absorbance detection limits.
Blank with the same buffer solution used to resuspend the analyte of interest. The
blanking solution should be a similar pH and ionic strength as the analyte solution.
•Run a blanking cycle to assess the absorbance contribution of your buffer solution. If the
buffer exhibits strong absorbance at or near an analysis wavelength, you may need to
choose a different buffer or application. See Choosing and Measuring a Blank for more
information.
For micro-volume measurements:
Ensure pedestal surfaces are properly cleaned and conditioned.
Ensure samples are homogeneous before taking a measurement. Avoid introducing
bubbles when mixing and pipetting.
Follow best practices for micro-volume measurements.
Use a 1-2 μL sample volume. See Recommended Sample Volumes for more
information.
For cuvette measurements (NanoDrop One
C
instruments only), use compatible cuvettes
and follow best practices for cuvette measurements.
Related Topics
Measure a Micro-Volume Sample
Measure a Sample Using a Cuvette
Best Practices for Micro-Volume Measurements
Best Practices for Cuvette Measurements
Prepare Samples and Blanks
Basic Instrument Operations