Measure Protein A280
Thermo Scientific NanoDrop One User Guide 63
–if Molar Extinction is selected
– enter molar extinction coefficient (
) in M
-1
cm
-1
divided by 1000 (i.e.,
/1000). For example, for protein with molar extinction coefficient of
210,000 M
-1
cm
-1
, enter 210
– enter molecular weight (MW) in kiloDaltons (kDa)
–tap OK to close New Protein Type box
After you choose OK, the new custom protein appears in the Type list in Protein A280
Setup and Proteins & Labels Setup.
Edit custom protein
– in Protein Editor, tap to select custom protein
– tap to show Edit Protein Type box
– edit any entries or settings
–tap OK
Tap a field to show keyboard;
to close, tap Done key