Measure Protein BCA
92 NanoDrop One User Guide Thermo Scientific
After the minimum number of standards has been measured for the selected curve type, a
message similar to the following appears:
Load more standards: returns to the setup screen where you can add or edit the
concentration value for any standard and then measure the standard.
Run samples: continues to sample measurement screen, after which standards can no
longer be edited.
• You can add, edit or delete a standard any time before the first sample measurement.
Add standard
– from standards measurement screen, tap > [application name] Setup
– tap the next empty Concentration field and enter the concentration value for the new
standard
–tap Done
Edit standard
– from standards measurement screen, tap > [application name] Setup
– tap the Concentration field and edit the concentration value
–tap Done
Delete standard
– from standards measurement screen, standard curve screen, or standards data table,
press and hold the row to show Standard Details box
–tap
The standard no longer appears in the table on the measurement screen and its
concentration value no longer appears on the setup screen.
Note You can use this method to delete the reference measurement; however, a new
reference must be measured immediately afterwards.