Thermo Scientific Orbitrap Fusion Series Hardware Manual 11
2
Scan Parameters
This chapter describes some of the scan parameter settings that you set in the Orbitrap Fusion
Series Tune application.
Scan Types
The following scan types are available:
MS Scan
MS
2
Scan and MS
n
Scan
SIM Scan
MS Scan
The full-scan MS (or MS
1
) scan type corresponds to a single stage of mass analysis—that is, a
scan power of n =1.
With the single-stage full scan, the ions formed in the API source are stored in the ion-routing
multipole (IRM). The mass spectrometer then transfers these ions to either the Orbitrap™ or
linear ion trap (LIT) to produce a full mass spectrum of the observable ions in the specified
mass range at a specific time point in the analysis.
Contents
Scan Types
Scan Rates for the Ion Trap Detector
Scan Mass-To-Charge Ratio Ranges
Resolutions for the Orbitrap Detector
Data Types
Ion Polarity Modes