Thermo Scientific Orbitrap Fusion Series Hardware Manual 49
System Shutdown, Startup, and Reset
When you are not using the Orbitrap Fusion Series system for short periods of time, place the
mass spectrometer in standby mode. For longer periods, for example, two or more months,
you can shut it down completely. In addition, many maintenance procedures for the
Orbitrap Fusion Series system require shutting down the mass spectrometer completely.
Shutting Down the System in an Emergency
Contents
• Shutting Down the System in an Emergency
• Placing the Mass Spectrometer in Standby Mode
• Turning On the Mass Spectrometer
• Shutting Down the Mass Spectrometer Completely
• Starting the System after a Complete Shutdown
• Resetting the Mass Spectrometer
• Resetting Calibration Parameters
• Restarting the Data System
• On/Off Status for MS Components Under Varying Power Conditions
CAUTION If you must turn off the mass spectrometer in an emergency, turn off the main
power switch located on the right-side power panel (Figure 3). This switch turns off all
power to the mass spectrometer, including the forepump, without harming components
within the instrument. However, do not use this method as part of the standard shutdown
procedure. Instead, see “Shutting Down the Mass Spectrometer Completely.”
To turn off the LC, autosampler, and data system computer in an emergency, use their
respective on/off switch or button.