Thermo Scientific Orbitrap Fusion Series Hardware Manual 67
8
Maintenance
This chapter provides routine maintenance procedures that you must perform to ensure
optimum performance of the Orbitrap Fusion Series MS. Optimum performance depends on
the maintenance of all parts of the instrument. You are responsible for maintaining the system
properly by performing the system maintenance procedures on a regular basis.
For a list of replaceable parts, see Chapter 9, “Replaceable Parts.
Note The following components are slightly different between the Orbitrap Fusion MS
and the Orbitrap Fusion Lumos MS: sweep cone, ion transfer tube, API source interface,
RF lens, MP00 RF lens, lens L0. Unless otherwise noted, use the Orbitrap Fusion MS
procedures.
CAUTION Heavy object. The Orbitrap Fusion Series MS, excluding its workbench,
weighs over 227 kg (500 lb). Never try to detach and move the instrument from its
workbench; you can suffer personal injury or damage the instrument. For additional
information, contact your local Thermo Fisher Scientific field service engineer.
Contents
Maintenance Schedule
Guidelines
Tools and Supplies
Maintaining the API Source Housing
Maintaining the API Source Interface
Maintaining the Forepump
Maintaining the Air Filter