8
Maintenance
Maintaining the API Source Interface
76 Orbitrap Fusion Series Hardware Manual Thermo Scientific
To clean the spray cone and O-ring
1. Soak the lint-free tissues or chamois-tipped swabs in a 50:50 solution of methanol/water,
and then clean the exterior surface of the spray cone.
2. (Orbitrap Fusion MS) Remove and inspect the O-ring located in the spray cone under
the entrance end of the ion transfer tube (Figure 41).
Figure 41. Spray cone, O-ring, ion transfer tube, and ion sweep cone (Orbitrap Fusion MS)
3. (Orbitrap Fusion MS) Clean the O-ring with methanol or replace it if necessary.
4. Using a magnification device, inspect the components for any residual lint or particulates.
5. (Orbitrap Fusion MS) Reinstall the O-ring in the spray cone.
To clean the ion transfer tube
1. If there is extreme contamination, follow these steps. If not, start with step 2.
a. Overnight, sonicate the component in a 10% solution of Liquinox in water.
b. Rinse the component with water, and then for 2 minutes force a strong stream of
water through the orifice.
c. For 30 minutes, sonicate the component in water.
Note The Orbitrap Fusion Lumos MS does not have an O-ring behind its ion
transfer tube.
Note Inspect the inside surfaces and edges to confirm that no lint or particulates are
present. Use plastic tweezers or a similar tool to remove the lint or particulate.
IMPORTANT Always use LC/MS-grade methanol and LC/MS-grade water.
Spray cone
Ion transfer tube
API cone seal
Gas inlet on the ion
sweep cone
Vespel™ O-ring