Maintaining the API Source Interface
78 Orbitrap Fusion Series Hardware Manual Thermo Scientific
Removing the API Source Interface
To remove the API source interface
1. Shut down and vent the system, and let it cool to room temperature.
Venting the mass spectrometer can take several minutes.
2. Unplug the mass spectrometer’s power supply cord from the electrical outlet.
3. Remove the API source housing.
4. Lift up the release latch, grasp the API source interface with your fingers, and then
carefully pull it out of the vacuum manifold (Figure 42).
Figure 42. API source interface removed from the vacuum manifold (Orbitrap Fusion MS)
CAUTION To avoid an electrical shock, be sure to follow the instructions in “Shutting
Down the Mass Spectrometer Completelyon page 52 before continuing with this
procedure.
CAUTION Hot surface. Allow heated components to cool to room temperature
(approximately 20 minutes) before you touch or service them.
CAUTION Do not disconnect the power supply cord at the mass spectrometer while
the other end is still plugged into the electrical outlet.
Release latch on the
API source interface
Vacuum manifold
Viton™ O-ring