Thermo Electron CorporationIX-4 Orion Star Series Meter
Conductivity Technique
Conductivity Setup
Description Display Range of Values Default
Value
Method
Specific
Conductivity Temperature
Compensation Selection
COnd
tC
LIn
OFF, LIn, nLF
Selects Temp Comp OFF,
Linear, or Ultra-pure Water
Lin
Yes
Conductivity Linear
Temperature Compensation
Coefficient Setting
COnd
COEF
2.1
0.0 - 10.0
Linear Temp Compensation
Coefficient in %/C
2.1
Yes
Conductivity TDS Factor
Setting
COnd
tdSF
0.49
0.00 - 10.0
TDS factor
0.49
Yes
Conductivity Auto-
Calibration Default Cell
Constant Setting
COnd
CELL
0.475
0.001 - 199.0
Cell Constant used for
Cond Auto-Standard
0.475
Yes
Conductivity Temperature
Reference Selection
COnd
trEF
25
15, 20, 25
Conductivity Reference
Temperature
25
Yes
Conductivity Cell Type &
Manual Ranging Selection
COnd
tyPE
PLnr
PLnr, Std, 1, 2, 3, 4, 5, 6, 7
Standard, Planer Cond Cell
Or sets Manual range 1-7
Std
Yes
1. In the measurement mode press
.
2. Press
/ to scroll through the Setup Menu’s until is displayed on
the Top line.
3. Press
to accept the selection and move the arrow icon to the middle line.