XIII-3
Orion Star Series Meter
Thermo Electron Corporation
Care and Maintenance
pH Events/Errors
Error
Code
Description
Press
to Clear
Cause Trouble Shooting
105 pH Auto-Buffer
Failure
Yes Auto-buffer not
recognized, will only
happen in remote
control
Check that electrode is in
correct buffer.
Check that correct buffer set in
selected in the method setup.
Clean electrode & refill
reference electrode
Perform a manual calibration.
Some electrodes may operate
outside of the range for pH
autocalibration. This may be
a sign of electrode aging and
is normal.
107 pH Calibration
Standard Error
Yes The milivolts being
measured are the
same for two different
standards.
Check that the two different
buffers are being used and
that the electrode is in the
correct buffer.
Make fresh buffers and repeat
the calibration.
109 Bad pH Slope Yes The pH slope is not in
the range of 80% to
120%
Repeat calibration using fresh
buffers
Clean electrodes and refill
reference electrode
Refer to electrode instruction
on how to check electrode