XIII-5
Orion Star Series Meter
Thermo Electron Corporation
Care and Maintenance
Conductivity Events/Errors
Error
Code
Description
Press
to Clear
Cause Trouble Shooting
705 Conductivity
Auto-Standard
Failure
Yes Auto-standard
not recognized,
will only happen
in remote control
Check that electrode is in correct
standard.
Recalibrate using fresh standards
Check that the appropriate cell constant
is entered in the method setup.
Clean electrode
Replatinize conductivity cell (if
applicable)
707 Conductivity
Calibration
Standard Error
Yes The conductance
being measured
is the same for
two different
standards.
Check that the two different standards
are being used and that the electrode is
in the correct standard.
Make fresh standards and repeat the
calibration.
709 Conductivity
Cell Constant
Yes The cell constant
is not in the
range of 0.001 to
199.0
Repeat calibration using fresh
standards
Clean electrodes
Replatinize conductivity cell (if
applicable
DO Events/Errors
Error
Code
Description
Press
to Clear
Cause Trouble Shooting
809 Bad DO Slope Yes The DO slope is
not in the range
of 3.92 to 19.7
nano-amps /
%saturation
Repeat calibration
On Air calibration make sure sponge in
calibration sleeve is damp.
On water calibration make sure water
is saturated with air by bubbling but
stirred to prevent keep bubbles off the
membrane surface.
Polish the cathode and replace the
membrane cap and fill solution.