Serial Communications 3104
REC XXXXX 3-6
- COMM. A SCROLL 2 -
Set parity port #1
> No parity <
CHOICE ENTER
- COMM. A SCROLL 3 -
Stop bits port #1
> 1 <
CHOICE ENTER
- COMM. A SCROLL 4 -
Wordlength port #1
> 8 <
CHOICE ENTER
- COMM. A SCROLL 5 -
Protocol port #1
> MODBUS <
CHOICE ENTER
Password: Service
Default: NO PARITY
Selections: EVEN PARITY, ODD PARITY, NO PARITY
Password: Service
Default: 1
Selections: 1,2
Password: Service
Default: 8
Selections: 7,8
The next screen defines the port use. Some commonly used protocols are
implemented in the system. Possible selections are:
PC-MASTER Thermo Ramsey proprietary protocol: Multidrop, Master Slave.
SIEMENS 3964R A proprietary protocol of Siemens. Point to point, Multi Master.
ALLEN-BRADLEY A proprietary protocol of Allen-Bradley.
DF1 Multidrop, Master Slave.
MODBUS A proprietary protocol of AEG. Multidrop, Master Slave.
PRINTER Not a protocol, selects printer output.
Password: Service
Default: MODBUS
Selections: PC-MASTER, SIEMENS 3964R, ALLEN-BRADLEY DF1, MODBUS, PRINTER