Do you have a question about the Thermotron S Series and is the answer not in the manual?
Review of DANGER, WARNING, and CAUTION labels on the chamber.
Essential guidelines to follow before operating the test chamber.
Critical warnings regarding ventilation, gas systems, and flammable materials.
Table detailing water consumption rates for different chamber models.
Requirements for water resistance (ohm-cm) and cartridge capacity.
Specifications for maximum floor, shelf, and total load capacity.
Temperature range, uniformity, and control specifications.
Humidity range, dry bulb, dew point, and control specifications.
Cooling and heating times for various temperature changes.
Overview of the chamber's electrical compartment and control transformer.
Description of air circulation, heating, cooling, and auxiliary cooling systems.
Operation of the LIGHT button on the 8200 display module.
How PTC controls product temperature versus air temperature.
Configuration and operation of the humidity mode.
Description of channels 1, 2, and 3 for temperature and humidity sensing.
How monitor channels provide additional temperature readings.
Details on Network (TCP/IP) and Computer I/O communication.
Function of the low limit switch for auxiliary cooling.
Overview of installation steps and applicable instructions.
Warnings about high voltages, moving parts, and gas systems.
Cautions regarding equipment handling, electrical grounding, and compressor use.
Guidelines for choosing a suitable installation location for the chamber.
Step-by-step instructions for positioning and leveling the chamber.
Procedure for installing and removing port gloves to maintain seal.
Steps for directly connecting demineralized water to the humidity system.
Procedure for installing the water recirculation system.
Steps for connecting the dry air purge system.
Method for properly installing port plugs for low humidity performance.
Connecting the condenser water supply and drain lines.
Guidelines for qualified technicians installing refrigeration lines.
Procedures for connecting main power wiring to the chamber.
Overview of the operation instructions and checked items.
Warnings for safe operation, avoiding hazards, and product protection.
Cautions regarding compressor handling, operating limits, and product protection devices.
Instructions for setting up and running tests using the programmer/controller.
Instructions for operating the chamber manually via the controller.
How to connect and disconnect the main power supply to the chamber.
Proper use of port gloves for handling products inside the chamber.
Guidance on identifying and resetting safety devices when the chamber fails.
Overview of preventive maintenance tasks and recommended frequency.
Warnings related to high voltages and performing maintenance.
Cautions regarding power disconnection and equipment handling during maintenance.
Procedure for checking refrigerant levels in R-23 or R-508B systems.
Procedure for checking refrigerant levels in R-404A, R-449A, R-507 systems.
Steps for cleaning the condenser fins to ensure proper operation.
Procedure for checking and replacing the demineralizer cartridge.
Guidelines for cleaning the chamber interior to prevent corrosion.
Procedure for changing the water filter to extend demineralizer life.
Steps for safely draining the vapor generator.
Procedure for checking and replacing the dry air purge filter.
Routine checks including operating currents, connections, and door seals.
Procedure for checking the oil level in compatible compressor types.
Method for testing various safety switches to ensure proper function.
Detailed procedure for checking refrigerant charge using a gauge manifold.
Procedure for checking discharge pressure for specific refrigerants.
Overview of common service procedures and professional help contacts.
Warnings about high voltages, working with refrigerants, and qualified personnel.
Cautions regarding electrical grounding, compressor damage, and power disconnection.
Procedure for adding oil to specific types of compressors.
Key points for liquid charging R-404A, R-449A, and R-507 refrigerants.
Procedure for collecting refrigerant as required by the Clean Air Act.
Detailed steps for charging the R-23 or R-508B refrigeration system.
Steps for charging R-404A, R-449A, or R-507 systems during a pull-down.
List of high usage, low usage, and other recommended spare parts.
Information required when contacting Thermotron for support.
Procedure for returning parts or instruments to Thermotron for repair/replacement.
Guidance on contacting manufacturers for non-Thermotron instrument issues.
Definitions of terms from Ambient to Port Plug.
Definitions of terms from Product to Transducer.
| Temperature Range | -70°C to +180°C |
|---|---|
| Humidity Range | 10% to 98% RH |
| Temperature Accuracy | ±0.3°C |
| Temperature Uniformity | ±0.5°C |
| Humidity Uniformity | ±2.0% RH |
| Power Supply | 208/230V |
| Humidity Accuracy | ±1.0% RH |