What to do if Thoratec CentriMag motor overheats?
- CChristina WolfAug 4, 2025
If the MOTOR OVER TEMP alert message appears, indicating the motor is overheating, clamp the return tubing and switch to the backup system.
What to do if Thoratec CentriMag motor overheats?
If the MOTOR OVER TEMP alert message appears, indicating the motor is overheating, clamp the return tubing and switch to the backup system.
What to do if Thoratec Medical Equipment main console or motor ceases to function?
If the main console or motor stops working, replace the hardware. Disconnect the blood pump from the main motor and console, then switch to a backup motor and console.
What to do if Thoratec Medical Equipment blood pump has been off for more than five minutes?
If the blood pump has been off for more than five minutes without adequate anticoagulation, replace the blood pump and other circuit components.
Pump Type | Centrifugal |
---|---|
Device Type | Centrifugal blood pump |
Indication | Short-term circulatory support for patients in need of hemodynamic support. |
Intended Use | Temporary circulatory support |
Control System | CentriMag Console |
Storage Temperature | -20°C to 60°C (-4°F to 140°F) |
Describes warnings and cautions to be considered when using the 2nd Generation CentriMag Primary Console.
Describes the procedure for unpacking the 2nd Generation CentriMag Primary Console and configuring it for use.
Describes how to operate the 2nd Generation CentriMag Primary Console.
Describes procedures for managing the 2nd Generation CentriMag Primary Console during defibrillation and in the event of equipment malfunction.
Provides critical warnings regarding system operation, compatibility, and backup requirements.
Describes the primary console, its front and back panels, and its capabilities.
Provides technical specifications for the 2nd Generation CentriMag Primary Console.
Details the layout and function of the primary console's operator controls and display.
Explains the controls and interface features of the Mag Monitor.
Guides the user through connecting power and turning on the system.
Describes the automatic self-tests performed upon system startup.
Explains how to configure console settings via the MENU keypad.
Instructions for setting the maximum flow alert threshold.
Instructions for setting the minimum flow alert threshold.
Steps to configure pressure alerts for the P1 transducer.
Guides on installing probes and activating pressure monitoring.
Describes how to start, stop, and adjust the blood pump speed.
Key operation step: step-by-step guide to starting the blood pump operation.
Explains the system's strategy for handling alarms and alerts.
Important for battery health: details the procedure for performing battery maintenance every 6 months.
Detailed procedure: step-by-step guide for the battery maintenance procedure.
Critical emergency procedure: instructions for replacing the main console/motor with backup units.