13
ENGLISH
BD-4600
To erase all data from one memory zone:
Press the MEMORY button for 4 ~ 6 seconds.
Press the START/STOP button to confirm
13 Symbol indication / Troubleshooting
Display symbol Condition/Cause Corrective action
The flashing heart will appear when the
pulse is detected during a measurement.
Measurements in process.
Do not talk or move.
Appears when the battery voltage is
excessively low or the position of batteries
is incorrect.
Replace all batteries with new ones.
Make sure the +/- polarity is correct.
IHB
Appears for 1 minute when the user was
talking, moving or shaking or an irregular
heart beat was detected during
measurement.
Repeat the measurement.
Note that you be relaxed for at least 5
minutes and sit comfortably and
quietly before you restart a
measurement.
Pressure system is unstable before
measurement
Fail to detect systolic pressure
Fail to detect diastolic pressure
Wrap the cuff correctly and tightly.
Ensure the connection is close
between the air plug and the unit.
Inflate over again after ensuring.
When measurement fails. Do not move your arm and body,
and keep quiet.
Measure over again according to
correct way.