EasyManua.ls Logo

Toro Greensmaster TriFlex 3400 - Page 211

Toro Greensmaster TriFlex 3400
332 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Greensmaster 3300/3400 Page 6 -- 41 Electrical System
NOTE: Solenoid coil resistance should be measured
withsolenoidatapproximately68
o
F(20
o
C).Resistance
maybe slightlydifferentthanlistedat differenttempera-
tures. Typically, a failed solenoid coil will either be
shorted (very low or no r esistance) or open (infinite re-
sistance).
4. Using a multimeter (ohms setting), measure resis-
tance between the two (2) connector terminals on the
solenoidvalvecoil.Theresistance forthes olenoidcoils
is identified below:
Solenoid Valve Coil
Resistance
PRV (mow) 7.1 ohms
S1, S3 and S4 (lift) 8.8 ohms
S2 (lift) 7.1 ohms
Leak Detector 7.2 ohms
5. If solenoid coil resistance is incorrect, replace sole-
noid (refer to appropriate control manifold service pro-
cedures in the Service and Repairs section of Chapter
5 -- Hydraulic System).
NOTE: To assist in troubleshooting, identical solenoid
valvecoilscanbeexchanged.Iftheproblemfollowsthe
exchanged coil, an electrical problem likely exists with
the coil. If the problem remains unchanged, something
other than the s olenoid coil is the problem source (e.g.
switch,circuitwiring,hydraulicproblem).Themowman-
ifoldPRVcoilisthesameastheliftmanifoldS2coil.The
lift manifold S1, S3 and S4 coils are identical.
6. After coiltestingiscompleted,connect wireharness
electrical connector to the solenoid valve coil.
7. Secure all machine components that were removed
to access hydraulic manifold.
1. Leak detector manifold 2. Solenoid valve
Figure 51
2
1
Electrical
System

Table of Contents

Other manuals for Toro Greensmaster TriFlex 3400

Related product manuals