The AC charger may be defective. Check voltage output of
charge with a voltmeter or use the DC charger to charge the
batteries.
Satisfactory Counts, but Results are in Error
Ensure that the handle is seated in bottom of notch while
taking readings.
Ensure the measurement depth on the display screen
corresponds to the actual source rod handle depth.
Check to see if an offset (density, trench, or moisture probe)
is enabled.
Ensure that the standard counts are correct.
Check calibration constants.
CAUTION
All information stored in the gauge except the
calibration constants is lost when the batteries are
disconnected.