Error Message Possible Causes Solution
depth mode. Contact 877-
TROXLER and press “2” and
TroxTechSupport@troxlerlab
s.com
initialize. Another
gauge or nuclear
source nearby.
Vertical surface too
close. Must be at
least 6 feet away.
Insufficient warm-up
This occurs if the source
rod is not in the SAFE
position during
initialization. If the error
continues to show in the
SAFE position, contact the
Troxler Service Center.
Try to store the data again.
Not Valid
Operator has tried to run
a drift test without
having a valid stat test.
then run a drift test.
Gauge Requires
Re-calibration!
The calibration constants
stored in the gauge have
become corrupted.
Contact the Troxler Service
Center.
Support
Code
invalid access code.
code (see the front of this
manual).
The source rod is not at a
valid depth when starting
a measurement.
position is at a valid depth
when starting a
measurement. Contact the