ICO 6-5
WARNING
Always wear heat-resistant gloves when handling any
hot substance.
Troxler recommends wearing a face shield when the
oven door is open and when handling the sample.
Use the basket carrier to move hot sample baskets
(see Figure 2-1 on page 2-8). Always keep the basket
carrier to the left side of your body, with the handle
between your body and the open oven door. This
helps prevent the operator from backing into the hot
insulation of the inside surface of the door.
7. Place the sample basket cover over the top sample
basket.
8. Use the lock arms to secure the sample basket cover to
the top basket.
9. Place the sample with the baskets, catch pan, and sample
basket cover in the center of the external scale. Record
the mass (W
T) from the external scale.
10. Determine the sample mass (W
S = WT – WP).
11. Set up the Auto-Print function for a printout that will
include %Loss, %AC, sample mass, and so on. Refer to
the auto-print section on page 5-10 for setup
instructions.
12. Press 〈START〉.
13. 15. The oven requests the sample mass (WS = WT – WP).
Enter the sample mass and press (ENTER).