4. Remove all weight from the platform.
5. Once the stable indicator is shown on the display,
press the ZERO key
Zero
. Digits on the display will
start blinking.
6. Use the numeric keys to enter the calibration weight
you are using.
7. Press the ZERO key
Zero
. LoAd will be shown on the
display.
8. Place the calibration weight on the platform.
9. Once the stable indicator is shown on the display,
press the ZERO key
Zero
. SUCCESS will be shown on
the display.
10. Scale will enter a self-test mode. Remove weight
before self-test mode is complete. After completing
the self-test, the scale will enter the weighing mode.
MULTI-POINT CALIBRATION (LINEAR)
Multi-point calibration uses four calibration points: zero,
1/3, 2/3 and full weight capacity of the scale. Refer to
the calibration weights charts below.
CALIBRATION WEIGHT (lb.)
Model # 1/3 WEIGHT 2/3 WEIGHT FULL CAPACITY
H-5819 2 lb. 7 lb. 11 lb.
H-5820 7 lb. 15 lb. 27 lb.
H-5821 15 lb. 30 lb. 55 lb.
H-5822 30 lb. 55 lb. 110 l b.
CALIBRATION WEIGHT (kg.)
Model # 1/3 WEIGHT 2/3 WEIGHT FULL CAPACITY
H-5819 1 kg. 3 kg. 5 kg.
H-5820 3 kg. 6 kg. 12 kg.
H-5821 6 kg. 14 kg. 25 kg.
H-5822 14 kg. 25 kg. 50 kg.
NOTE: Calibration weights are sold separately.
See uline.com for details.
1. Turn on the scale and press the TARE key
Tare
during
the self-test. PO CAL will be shown on the display.
2. To select multi-point calibration, press the TARE key .
P 1 LCA will be shown on the display.
3. Press the ZERO key
Zero
. Lb will be shown on the
display.
a. If calibrating in lb., continue to step 4.
b. If calibrating in kg., press the TARE key
Tare
.
G will be shown on the display.
4. Press the ZERO key
Zero
. LoAd0 will be shown on the
display.
5. Remove all weight from the platform.
6. Once the stable indicator is shown on the display,
press the ZERO key
Zero
. LoAd1 will be shown on the
display.
7. Place 1/3 of the scale capacity on the platform.
8. Once the stable indicator is shown on the display,
press the ZERO key
Zero
. LoAd2 will be shown on the
display.
9. Place 2/3 of the scale capacity on the platform.
10. Once the stable indicator is shown on the display,
press the ZERO key
Zero
. LoAd3 will be shown on the
display.
11. Place the full scale capacity on the platform.
12. Once the stable indicator is shown on the display,
press the ZERO key
Zero
. SUCCESS will be shown on
the display.
13. Scale will enter a self-test mode. Remove weight before
self-test mode is complete. After completing the
self-test, the scale will enter the weighing mode.
OPERATION CONTINUED
LoAd
SUCCES
LoAd0
LoAd1
PO CAL
LoAd2
LoAd3
SUCCES
P I LCA
Tare