PAGE 6 OF 36 0319 IH-5836
ZERO
If there is a minor weight displayed without anything
on the platform, press the ZERO key
to clear the
display. Zero indicator will be shown on the display.
TARE
When weighing an item that is in a container, taring
stores the container weight to memory so only the item
weight is displayed.
To store tare weight into memory and set display to zero:
1. Place an empty container on the platform.
2. Press the TARE key
. The net indicator will be
displayed on the bottom of the display.
3. The container's weight is then stored in the scale's
memory and zero is displayed.
4. Add pieces to the container. As the pieces are
added, their net weight will be displayed.
5. Press the G/N key
to display the gross weight
and net weight.
6. Removing the container from the platform will cause
the scale to display the container's negative weight.
Pressing the TARE key
will readjust the scale to zero.
WEIGHING
Use this mode to weigh items.
1. Place item to be weighed on the scale platform.
Weight will be show in the display.
COUNTING
Use this mode to count parts of uniform weight.
SAMPLE OF UNIT WEIGHT
The scale cannot determine how many pieces are on a
platform without a sample.
1. Place objects intended for sampling on the platform.
2. Press the PCS/SET key to enter the counting
mode. P 10 will be shown on the display.
3. Press the TARE key
to change the sample
quantity – 10, 20, 50, 100 or 200 units.
P 10 / P 20 / P 50 / P 100 / P 200
4. Press the ZERO key . "----" will be shown on the
display followed by the sample quantity.
5. Place remaining objects on the platform. The total
quantity will be displayed.
6. Press the PCS/SET key
to exit to the normal
weighing mode.
LACK OF UNIT WEIGHT
The minimum weight must be at least 20 times the scale
accuracy for accurate calculations. Reference the
table below for minimum weights.
Model #
Minimum Piece
Weight
Minimum Sample
Weight
H-5836 0.01 lb. 0.4 lb.
H-5837 0.025 lb. 1 lb.
If the unit weight is too light, another scale is needed.
Continuing to use this scale will result in less accurate
results.
CHECK WEIGHING
Use this mode to set a number for a weighing check. When
the weight of the objects on the platform is between the
preset checking numbers, the alarm sound beeps repeatedly.
SET CHECK WEIGHT
NOTE: Weight on platform must be equal to
or greater than minimum weight in order for
check weighing mode to operate accurately.
1. Press and release the UNIT key
and the
PCS/SET key simultaneously. SEt H will be shown
on the display.
2. To set the high limit, press the ZERO key
.
00000 will be shown on the display with the last
digit blinking.
3. Key in the high limit. Use the G/N key
and the
ACC/TTL key
to move the active digit left and
right.
Use the TARE key to increase the numeric value.
4. Press the ZERO key
to confirm. Set H will be
shown on the display.
5. Press the TARE key
until Set L is shown on the
display.
6. To set the low limit, press the ZERO key
. 00000
will be shown on the display with the last digit blinking.
OPERATION CONTINUED