Type | Temperature and Humidity Data Logger |
---|---|
Humidity Range | 0 to 100 %RH |
Memory Capacity | 16, 000 readings |
Data Points | 16, 000 |
Sampling Rate | 1 second to 24 hours |
PC Interface | USB |
Temperature Range | -40 to 70 °C |
Temperature Accuracy | ±0.5°C (±0.9°F) |
Humidity Accuracy | ±3 %RH |
Resolution | 0.1 °C and 0.1 %RH |
Battery Type | Lithium Battery |
Software | Included |
Lists key capabilities like IP67 rating, multi-parameter display, and thermocouple compatibility.
Highlights common safety practices for working with temperature-critical environments and the device.
Identifies the four thermocouple input terminals (T1, T2, T3, T4) on the device.
Refers to the device's primary display screen for showing readings and status.
Details the various buttons on the device used for operation and navigation.
Locates the USB port for PC connectivity and data transfer.
Details the main display area showing current temperature readings for selected channels.
Explains various status icons like SETUP, RECALL, OFFSET, HOLD, USB connection, and logging status.
Indicates the units (°C, °F, K) for temperature display.
Shows secondary information like internal temp, memory status, MIN/MAX/AVG readings.
Displays the 24-hour clock, interval length, or elapsed time during logging.
Controls the backlight on/off and powers the thermometer on/off with a long press.
Toggles between T1/T2 and T3/T4 or scrolls to temperature differential.
Steps through MIN, MAX, or AVG readings; long press turns off display.
Freezes readings or switches between Celsius, Fahrenheit, and Kelvin with a long press.
Covers connecting thermocouples and powering on the device before configuration.
Explains how to enter, exit, and navigate the device's setup menu for configuration.
Procedure for setting the current date on the thermometer.
How to select preferred units for time display (e.g., h:m).
Steps to set the current time on the thermometer, including 24-hour format.
Instructions for selecting the correct thermocouple type (e.g., KJET) for accurate readings.
Configures the automatic power-off feature to conserve battery life.
Sets the frequency for storing readings during data logging sessions.
Procedure to compensate for thermocouple errors by adjusting the reading offset.
Returns the thermometer to factory default settings, clearing custom configurations.
Instructions for properly connecting thermocouples, including color coding.
How to view temperature measurements, toggle between channels, and interpret display messages.
Freezes the current readings on the display for easier viewing or recording.
Accesses and cycles through minimum, maximum, and average recorded values.
Details the process of using the offset function in SETUP to correct for thermocouple measurement errors.
Explains how the thermometer stores logged readings in memory and the capacity.
Instructions for initiating and terminating data logging sessions.
Procedure to erase all stored logged readings from the device's memory.
How to access and review previously recorded data stored in memory.
Information on connecting the thermometer to a PC using WS600 software for data acquisition.
Emphasizes that repair and service should only be done by qualified personnel.
Provides tips for long-term, reliable service, including calibration and environmental care.
Instructions for cleaning the meter's case without damaging its components.
Steps for properly winding and storing thermocouples on the device.
Details operating temperature, storage temperature, humidity, and altitude limits.
Includes dimensions, weight, battery type, certification, and safety ratings.
Details the specifications for the K-Type bead thermocouple accessory.
Covers measurement ranges, display resolution, accuracy, and temperature coefficients.