Page 21
CONNECTORS
Connecting load cells to the indicator
The indicator can connect with 6 load cells of 350Ω at most
4 wire or 6 wire load cell connections are both okay
Please contact us directly if you have other special needs for your application
There are two connection methods between the load cell and indicator Quick Disconnect as
shown below:
Hardwire (Using Inner Terminal Block Connection:
Note: Make sure you follow all the anti-static rules to avoid damage to your indicator
Excitation voltage: 5V DC
Largest output current: 120 mA
Excitation circuit: 5 VDC, 4 wire connection, 6 load cell of 350ohm maximum
Open the back cover of the weighing indicator, and insert signal cable to the terminal
block; Make sure the screw on terminal block is xed tightly.