9
INDICATOR PARAMETER SETTINGS
This indicator has 5 different setting modes, [F 1], [F 2], [F 3], [F 4] & [F 5].
Each mode changes different indicator parameters.
To enter calibration/parameter settings, follow the procedure below:
1. Press and hold the SWITCH and PRINT key at the same time for 2 seconds
2. Navigate through the setting modes (F1 to F5) by using the arrow keys and return keys as
labeled under each indicator button (
, , , )
3. Press the PRINT key to enter/edit the parameter setting
4. Press the CHECK
key to save and exit settings at any time
F 1 Menu
Setting working mode, unit, decimal, graduation, and capacity
Step Parameter Display Settings/Options
Enter F1 Menu
F 1
1 Mode
Mode 2
Choose how many weighing pad’s you’ll be using
Options: 1/2/3/4
2 Unit
Unit-0
Choose default unit
0=kg
1=lb
3 Decimal setting
dot--2
0 = no decimal
1 = #.#
2 = #.##
3 = #.###
4 = #.####
4 Graduation Setting
(readability of the least
signicant digit)
E - 05
Options: 1/2/4/10/20/50
Example with no decimal places (ie step 3=0)
1 = 1 lb
2 = 2 lb
5 = 5 lb
10 = 10 lb
20 = 20 lb
50 = 50 lb
5 Max Capacity
Full-0
0 = do not change the max capacity
1= change the max capacity
ex. If step 2=0 and step 3=2
when changing max capacity 0100.00 = 100kg
Then press to save your settings