The Pt100 temperature sensor is soldered to the probe. Do not try to
remove it. Due to the operating principles of the sensor, no maintenance is required.
CAUTION!
1. Remove the probe from inside the radiation shield.
2. Carefully remove the filter from the probe. Holding the filter by the plastic part, rotate
the
filter counter-clockwise, then pull it out.
Do not touch the sensor heads.CAUTION!
1 Filter
2 O-ring
3 HUMICAP humidity sensor
4 Pt100 temperature sensor
3. After removing the filter, check the Oring and change it if necessary.
4. Without delay, install the new
filter carefully on the probe.
5. Install the probe back inside the radiation shield.