Replacing the INTERCAP® Humidity Sensor
n INTERCAP® sensor (Vaisala order code 15778HM)
n Flat head screwdriver (for opening the transmitter)
To avoid contaminating or damaging the sensor:
n Handle the sensor by its plastic frame.
n Do not touch the sensor surface in the middle.
n Wear lint-free ESDgloves.
1. Disconnect the transmitter body from the mounting base.
2. Locate the INTERCAP® sensor on the component board. See Figure 2 on
page8.
3. Disconnect the old INTERCAP® sensor by pulling it straight out of its
socket.
4. Insert the new INTERCAP® sensor in the socket.
5. Reconnect the transmitter to the mounting base.
6. Check the output of the transmitter (or the display if included on the model)
to verify that the humidity measurement is working normally and the
transmitter is not in the error state. For more information on the error state,
see Error State on page33.