Error State
If the transmitter detects a serious hardware or software error, or the measured
reading is well outside the scaled range, the analog outputs are set into a
defined error level instead of the measured result. The error level depends on
the output type:
n For 4 ... 20 mA output, the error level is 3.6 mA.
n For 0 ... 10 V output, the error level is 11 V.
If the cause of the error is removed, the transmitter resumes normal operation of
analog outputs. Some errors may require a reset of the transmitter to recover.
If the error state is caused by a measurement result being well outside the
scaled output range, only affected channels are set to the error state. For more
information, see Analog Output Overrange Behavior on page13.
Examples of situations that can cause the error state:
n The transmitter measures over 6000 ppm CO
2
.
n The CO
2
measurement module (GM10) is disconnected.
33