Command Description
ctclr
Clear user calibration for temperature measurement. Factory calibration
remains.
ctext
Show or set adjustment information text.
fcrh
Two-point calibration after humidity sensor change.
l
Show adjustment oset and gain.
li
Set adjustment oset and gain.
Other Commands
filt
Show or set measurement filtering.
frestore
Restore factory settings. Clears all user settings, factory calibration remains.
reset
Reset the device.
6.5 Device Information and Status
Table 6 ? Command
Syntax Description
?<cr> Show listing of device information.
??<cr> Show listing of device information even if device
is in poll mode and connection has not been
opened using the open command.
Chapter 6 – Using Service Port With Terminal Application
35