Syntax Description
Example:
?
HMDW110 / 1.0.6
Serial number : H0134007
Batch number : H0130002
Sensor number : H0090003
Sensor model : Humicap 180R
Order code : HMD1102A1VA1
Cal. date : 20140619
Cal. info : VAISALA/HEL
Time : 01:43:43
Serial mode : STOP
Baud P D S : 19200 N 8 1
Output interval: 5 S
Serial delay : 25
Address : 0
Filter : 1.000
Ch1 output : 4 ... 20 mA
Ch2 output : 4 ... 20 mA
Ch1 RH lo : 0.00 %RH
Ch1 RH hi : 100.00 %RH
Ch2 T lo : -40.00 'C
Ch2 T hi : 60.00 'C
Table 7 Errs Command
Syntax Description
errs<cr> Show active error(s). Possible error messages
are listed in Error Messages in Insight Software
(page 70).
Example (no active errors):
errs
0000h
No errors
Table 8 Error Messages on ERRS Command
ERRS Command Response Corresponding Error
T MEAS error
Temperature measurement error. [44]
F meas error
Humidity measurement error. [45]
RH sensor failure
Humidity sensor failure. [46]
Frequency measurement outside the
permissible value range
Capacitance reference error. [47]