Syntax Description
Example:
intv 5 s
Output interval: 5 S
6.6.2 Measurement Output Format
Table 16 Unit Command
Syntax Description
unit<cr> Show current setting of the unit command.
unit [m|n]<cr> Set types of units used on the serial line.
m = metric units, for example, Celsius
n = non-metric units, for example, Fahrenheit
Example (set units to non-metric):
unit n
Units : Non metric
Table 17 Form Command
Syntax Description
form<cr> Show the currently used measurement format.
form /<cr> Reset measurement format to default.
form [sss]<cr> Set a new measurement format.
sss = String consisting of modifiers and
abbreviations for measured parameters. See
Table 18 (page 42) and Table 19 (page 42).
Maximum length 127 characters.
Example: show currently used measurement format (default format shown):
form
3.1 "T=" T " " U3 3.1 "RH=" RH " " U4 3.1 "Td=" Td " " U3 3.1 "Tw=" Tw
" " U3 4.1 "h=" h " " U7 \r \n
Output example:
send
T= 25.1 'C RH= 39.4 %RH Td= 10.3 'C Tw= 16.2 'C h= 45.1 kJ/kg
Chapter 6 – Using Service Port With Terminal Application
41