Syntax Description
Example (set serial line delay to 200 milliseconds):
sdelay 50
Serial delay : 50
Table 25 Smode Command
Syntax Description
smode<cr> Show current start-up operating mode of the
serial line, and prompt to enter new mode. New
mode is taken into use when the device is reset
or powered up.
smode [mode]<cr> Set serial line start-up operating mode.
Available modes are:
stop = No automatic output. All commands
available. Default mode.
run = Automatic output of measurement
messages. You must stop output with the s
command before entering other commands.
poll = No automatic output. Will respond to
addressed send command and ?? command.
You can use other commands after opening a
connection using an addressed open command.
Modbus = No automatic output. Measurement
outputs must be read from the transmitter using
the Modbus protocol.
Example (set device to poll mode):
smode
Serial mode : STOP ? poll
More Information
‣
Modbus Communication (page 28)
‣
Modbus Registers (page 79)
Chapter 6 – Using Service Port With Terminal Application