6.8 Calibration and Adjustment Commands
Before using the calibration and adjustment commands, read
through Calibration and Adjustment (page 54).
CAUTION!
Table 30 Cdate Command
Syntax Description
cdate<cr> Show currently stored calibration date.
cdate [yyyymmdd]<cr> Set a new calibration date.
yyyymmdd = Date in format year (yyyy) month
(mm) date (dd).
Example (set calibration date to June 1st, 2018):
cdate 20180601
Cal. date : 20180601
Table 31 Crh Command
Syntax Description
crh<cr> Start the two-point humidity calibration and
adjustment sequence. For a full adjustment
procedure, see Two-Point Humidity Calibration
and Adjustment Using Computer and HMK15
Humidity Calibrator (page 59).
crh [ref]<cr> Perform a one-point adjustment at the current
relative humidity. The device calculates the
required correction based on the reference
humidity that you enter.
ref = the correct relative humidity at the
environment where the sensor is now.
Example :
crh
RH : 11.5379 1. ref ? 11.3
Press any key when ready ...
RH : 74.9684 2. ref ? 75.4
OK