Table 35 Ctext Command
Syntax Description
ctext<cr> Show currently stored calibration text.
ctext [sss]<cr> Set a new calibration text.
sss = Text string, maximum length 24 characters.
Example (set calibration text to Lab1/John):
ctext Lab1/John
Cal. info : Lab1/John
Table 36 Fcrh Command
Syntax Description
fcrh<cr> Start the two-point humidity calibration and
adjustment sequence. If you have changed the
humidity sensor of the device yourself, you must
perform a two-point humidity calibration and
adjustment using this command. Follow the
procedure in Two-Point Humidity Calibration
and Adjustment Using Computer and HMK15
Humidity Calibrator (page 59).
Example:
fcrh
RH : 11.3143 1. ref ? 11.3
Press any key when ready ...
RH : 75.0012 2. ref ? 75.4
OK
Table 37 L Command
Syntax Description
l<cr> Show the current oset and gain parameters for
user adjustment. This command is useful for
checking if user adjustments have been applied
for humidity and temperature measurement.
HMDW110 User Guide M211726EN-D
50