Table 35 Ctext Command
Syntax Description
ctext<cr> Show currently stored calibration text.
ctext [sss]<cr> Set a new calibration text.
sss = Text string, maximum length 24 characters.
Example (set calibration text to Lab1/John):
ctext Lab1/John
Cal. info : Lab1/John
Table 36 Fcrh Command
Syntax Description
fcrh<cr> Start the two-point humidity calibration and
adjustment sequence. If you have changed the
humidity sensor of the device yourself, you must
perform a two-point humidity calibration and
adjustment using this command. Follow the
procedure in Two-Point Humidity Calibration
and Adjustment Using Computer and HMK15
Humidity Calibrator (page 59).
Example:
fcrh
RH : 11.3143 1. ref ? 11.3
Press any key when ready ...
RH : 75.0012 2. ref ? 75.4
OK
Table 37 L Command
Syntax Description
l<cr> Show the current oset and gain parameters for
user adjustment. This command is useful for
checking if user adjustments have been applied
for humidity and temperature measurement.