6.9 Other Commands
Table 39 Filt Command
Syntax Description
filt [f.fff]<cr> Set the speed at which the latest measurement
result is integrated into the humidity and
temperature readings. The command aects
both analog output and serial line output.
f.f = Measurement filtering setting, range
0.001 ... 1.0.
1.0 = No filtering, latest measurement is
output without averaging
0.5 = Average of last two measurements
0.1 = Average of approximately 16
measurements
filt<cr> View the current setting. Also prompts you to
enter a new value.
Example (view the current value and set filtering to value 0.5):
filt
Filter : 1.000 ? 0.5
Table 40 Frestore Command
Syntax Description
frestore<cr> Restores the factory default settings. All user-
made settings are lost, including user
calibration. Reset the transmitter after giving
this command.
If you have replaced the HUMICAP
®
humidity
sensor of the device yourself, you must redo the
two-point humidity calibration using the FCRH
command after performing a factory reset.
Example:
frestore
Factory settings restored
reset
HMDW110 / 2.2.3
HMDW110 User Guide M211726EN-D
52