11. The adjustment mode now shows the value of the selected parameter from both
instruments, and a delta parameter (for example, ΔRH) that shows the dierence
between the two.
Wait at least 15 minutes for the measurement to stabilize, and for the value of the delta
parameter to stop changing. Stabilization may take a much longer time, depending on
factors such as temperature dierences and airflow around the sensors.
12. The dierence between the reading should be no greater than the combined
measurement uncertainty of the HMDW110 series transmitter and the reference probe.
Depending on the result, proceed as follows:
If the HMDW110 series transmitter is within its accuracy specification, there is no need
to proceed with the adjustment. Select Back > Exit to leave the adjustment mode.
If adjustment is required, continue with the next step.
13. Select Adjust > To same as RH (I)
14. The MI70 will ask Do you really want to adjust?. Select Yes.
MI70 shows the text Adjustment done and returns to the adjustment mode after a few
seconds.
15. Check the value of the delta parameter to verify that the adjustment has taken eect.
16. Select Back to return to the parameter selection screen.
17. Select a new parameter for adjustment and proceed as instructed above, or select Exit to
leave the adjustment mode.
18. Turn o the MI70 indicator and disconnect the connection cable from the service port.
19. Reconnect the terminal blocks and close the cover.