15. Give the l command to verify that the values of the adjustment parameters Cp oset and
Cp gain have changed.
Example: User adjustment parameters after two-point humidity adjustment (your values
will be dierent):
Cp offset : -2.21914053E-03
Cp gain : 1.00767076E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
16. Remove the probe from the salt jar and reconnect the filter. Plug the hole on the salt jar.
17. Update the calibration date and calibration information on the transmitter using the
cdate and ctext commands.
Example: Set calibration date to 2018-07-04 and calibration info text to "Lab2/Mike".
cdate 20180704
Cal. date : 20180704
ctext Lab2/Mike
Cal. info : Lab2/Mike
18. Disconnect the connection cable from the service port.
19. Reconnect the terminal blocks and close the cover.
7.6
Two-Point Temperature Calibration and
Adjustment Using Computer
You can also use Vaisala Insight software for performing calibration and
adjustment. See Vaisala Insight Software (page 29).