SERVICE
D+ D− GND
GND
+ VIN − + CH1 − + CH2 − + CH3 − PURGE
Figure 4 Connecting module to Insight
1. Open Insight software.
2. Connect the USB adapter to a free USB port on the PC or USB hub.
3. Connect the USB cable to the service port of the module. The USB cable powers up the
module.
5.3.2 Configuration options
Select > Configure Device to access configuration options in Insight software.
The available configuration options correspond to the Modbus configuration registers. Insight
software is the recommended way to change the device configuration.
You can restore the device back to its default settings using the Factory Default
Settings > Restore Settings function. Doing this will also clear any user
adjustment and restore the latest factory calibration.
More information
Modbus reference (page 32)
5.3.3 Configuring condensation prevention
1. Connect the module to Insight. See Connecting to Insight software (page 20).
2. Select > Configure Device.
3. Set the Condensation prevention on/o switch to the desired setting.
The switch is not available if the module does not support any condensation
prevention heating features, or if the probe firmware version is older than
1.2.0. Modules with older firmware versions will show configuration switches
for individual condensation prevention features as applicable.
Chapter 5 – Operation
21