EasyManua.ls Logo

Vaisala HUMICAP HMM170 - Dimensions

Vaisala HUMICAP HMM170
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
°C
0
0-80 -60 -40 -20 20 40 60 80 100 120 140 160
180
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.1
0.2
0.3
0.4
0.5
0.6
0.7
°C
Figure 5 Temperature measurement accuracy over temperature range
Table 8 HMM170 operating environment
Property Description/Value
Operating temperature for component board −40 … +60 °C (−40 … +140 °F)
Operating humidity for component board 0 … 100 %RH, non-condensing
Storage temperature −55 … +80 °C (−67 … +176 °F)
Operating pressure 0 … 10 bar
Table 9 HMM170 inputs and outputs
Property Description/Value
Three analog outputs (selectable and scalable) 0 – 20 mA, 4 – 20 mA
0 – 1 V, 0 – 5 V, 1 – 5 V, or 0 – 10 V
Typical accuracy of analog output at +20 °C
(+68 °F)
±0.05 % full scale
Typical temperature dependence of analog
output
0.005 %/˚C (0.003 %/°F) full scale
External load R
L
< 500 Ω
Digital output RS-485 serial, Modbus
Service port M8 connector for USB cable
Start-up time 3 s at power-up
Wire size
0.5 … 1.5 mm
2
(20 … 16 AWG)
Supply voltage