Chapter 3 _______________________________________________________________ Installation
VAISALA_______________________________________________________________________ 41
General Instructions for Probes with a
Cable
Mount the probes with a cable horizontally; this way, any water
condensing on the tube cannot flow onto the sensor.
0507-024
Figure 23 Horizontal Mounting of Probe
The following numbers refer to Figure 23 above:
1 = To be sealed
2 = To be insulated
3 = Insulate the cable
4 = Let the cable hang loosely. This prevents condensed water
running to the probe along the cable.